The reported event that could not be confirmed, based on available ct scans and medical expert assessment the device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed "ct-scan does not show clear signs of loosening, there are very small areas with minor radiolucency (< approx.1 mm).More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|