Model Number 1458Q/86
Device Problem Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
Further information requested but was not received.
 
Event Description
It was reported that the patient presented for the generator change procedure.Prior to the procedure, upon interrogation, the left ventricular (lv) lead exhibited a high capture threshold.The lv lead was explanted and replaced.
 
Manufacturer Narrative
The reported event of a high capture threshold was not confirmed.As received, a complete lead was returned in two pieces for analysis.Electrical and x-ray examination testing did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies except for procedural damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18196915
MDR Text Key328883022
Report Number2017865-2023-73099
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model Number1458Q/86
Device Lot NumberA000078063
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/02/2023
Initial Date FDA Received11/22/2023
Supplement Dates Manufacturer Received12/04/2023
Supplement Dates FDA Received12/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA LEAD; GALLANT ICD; TENDRIL STS LEAD
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexMale
-
-