• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR TIP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR TIP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71363077
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  malfunction  
Event Description
It was reported that, during a tka surgery, an r3 offset impactor tip broke on impaction.All pieces were retrieved by hand, nothing fell in the patient.It is unknown how the surgery was completed or if there was any delay.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Manufacturer Narrative
Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals the device broke into several pieces.All pieces were returned.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.Device batch number was not provided; thus, an evaluation of the manufacturing records could not be performed.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Manufacturer Narrative
H10; additional information.Updated section b3 and b5.
 
Event Description
It was reported that, during a tka surgery, a r3 offset impactor tip broke on impaction.All pieces were retrieved by hand, nothing fell in the patient.The surgery was completed, without any delay, with a s+n back-up device.No injuries were reported.
 
Manufacturer Narrative
H11: corrected data section d9 and h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 OFFSET IMPACTOR TIP
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18204474
MDR Text Key328909738
Report Number1020279-2023-02315
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71363077
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-