Model Number 71019
Device Problem Insufficient Information (3190)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/27/2023
Event Type  Injury  
Event Description
Patient was implanted with the nalu peripheral nerve stimulator system on (b)(6) 2023 targeting the cluneal nerve to treat bilateral hip pain.Approximately one week after implant the patient reported difficulty with communication between the implantable pulse generator (ipg) and the external therapy discs.A surgical revision was performed (b)(6) 2023 to move the ipg from the lateral flank to a more medial position in order to correct the communication issues.No components were explanted or replaced during the procedure.
 
Manufacturer Narrative
The patient successfully completed a trial phase and wear study prior to implanting the nalu system.Intra-op and post-op testing of the system showed good connectivity.No imaging was performed prior to moving the device to confirm the placement of the ipg.It is possible the ipg migrated after implant or that the placement of the ipg was in an area of loose tissue, causing interference in communication as the patient was moved out of the prone position.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key18207867
MDR Text Key328957990
Report Number3015425075-2023-00291
Device Sequence Number1
Product Code GZF
UDI-Device Identifier00812537036441
UDI-Public01008125370364411123071017260710
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71019
Device Catalogue Number71019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-