• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. UNSPECIFIED BD INFUSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. UNSPECIFIED BD INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number UNKNOWN
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter email address- (b)(6).H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that unspecified bd infusion set over infused.The following information was received by the initial reporter with the verbatim: (b)(6) 2023 - lvp giving frequent air-in-line alarms.Patient could visibly see bubbles in the line.Nurses were trying to get the bubbles out of the line and eventually told the patient "it's okay if a few bubbles get in there."(b)(6) 2023 - same pump as before.Continuing the frequent air-in-line alarms.Nurse told patient "when it alarms like that: just push the reset button."patient had to push the reset button about 40 times this day.Patient requests to have the lvp exchanged and the nurses change it out but the new lvp is also alarming.When these alarms get annoying for the nurse she removed the line from the pump and just let it hang.(b)(6) 2023 - alarms again on the lvp this day the line is back in the lvp.Patient is concerned about this day because the iv bag emptied in about 10 minues instead of the usual hour he was used to.Patient told the nurses he was sick of pushing the reset button and the nurses told him that unfortunately the equipment at this location isn't very well kept by the biomedical engineers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD INFUSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18207869
MDR Text Key328968552
Report Number9616066-2023-02263
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-