• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458QL/86
Device Problems Use of Device Problem (1670); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, the left ventricular (lv) lead got dislodged after initial implant.The physician made a second attempt to place the lead but was unsuccessful because it could not be fixed.The lv lead was then removed and replaced.The patient was in stable condition throughout.
 
Manufacturer Narrative
The reported event indicated unable to implant and lead dislodgement.Final analysis found that as received, a complete lead was returned in one piece.The s-curve hump height was measured within specification.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual examination found no anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18207882
MDR Text Key328966416
Report Number2017865-2023-93183
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510219
UDI-Public05414734510219
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458QL/86
Device Lot NumberA000140439
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received11/26/2023
Supplement Dates Manufacturer Received01/26/2024
Supplement Dates FDA Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-