Catalog Number SGC0702
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that a patient presented with grade 4 functional mitral regurgitation (mr) for a mitraclip procedure.A ntw clip could not be implanted due to gradient issues.It was decided to remove and replace the clip with an nt.While retracting the ntw clip into the steerable guide catheter (sgc), it got stuck at the hemostasis valve.The clip could not be removed due to opened/deformed clip arms.The sgc was promptly removed from the heart to prevent air embolism.The hemostasis valve broke from the guide handle while trying to remove the clip.The procedure was able to be completed with one clip implanted, and an mr grade reduction to <1.There were no adverse patient effects or clinically significant delay.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported difficulty removing the cds from the sgc (clip got stuck at the hemostasis valve).The reported broken hemostasis valve was due to troubleshooting maneuvers of the difficulty to remove the device as the user broke it while trying to remove the clip from the valve.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER (CE)
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18208730
MDR Text Key328996860
Report Number2135147-2023-05178
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSGC0702
Device Lot Number30725R1111
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MITRACLIP; STEERABLE GUIDE CATHETER
-
-