• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER Back to Search Results
Model Number U225
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Asystole (4442)
Event Date 05/12/2023
Event Type  malfunction  
Event Description
It was reported that the health care provider reached out to technical services (ts) to know if the electrocardiogram exhibited noise or ventricular tachycardia episode.Upon review, it was noted that this pacemaker triggered to lead safety switch suggesting high impedance out of range on the right ventricular (rv).Noise, oversensing, and pacing inhibition greater than 2 seconds were reported.Then the patient came into the clinic to troubleshoot, and it was noted that the patient had several computed tomography (ct).The sensitivity was adjusted and check the rv lead was recommended.Currently, this device remains in service and no additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18209125
MDR Text Key329003538
Report Number2124215-2023-66762
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559426
UDI-Public00802526559426
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/15/2023
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number720552
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received11/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
-
-