MAUDE Adverse Event Report: ABBOTT MEDICAL EPIC¿ SUPRA VALVE (AORTIC); HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EPIC¿ SUPRA VALVE (AORTIC); HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Catalog Number ESP200-21
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Obstruction/Occlusion (2422)
Event Date 10/18/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2023, a 21mm epic plus supra was chosen for implantation as a aortic valve replacement.The patient presented with lower coronary ostia height.The valve was implanted, but the stent post obstructed/impinged the coronary ostia, causing blood pressure to lower.The valve was explanted and a non-abbott 19mm valve was implanted at an intra-annular position successfully.Blood pressure returned to normal.The patient remained on cardio-pulmonary bypass throughout the procedure.There was a 40 minute delay to replace the valve, which resulted in no patient effects or intervention.The patient was reported to be stable.
 
Manufacturer Narrative
An event of explant due to lowering of blood pressure post implantation was reported.Information from the field indicated that the patient presented with lower coronary ostia height and the stent post obstructed/impinged the coronary ostia.The device was returned to abbott for investigation which revealed that the sewing cuff was intact and contained blood/bodily fluid.There were no perforations or tears observed and the cusps were mobile.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC¿ SUPRA VALVE (AORTIC)
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL LTDA. REG#3001883144
1301rua profvieira demendonça
bairro engenho nogueira 31.31 0-26
BR   31.310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18209166
MDR Text Key329004024
Report Number2135147-2023-05181
Device Sequence Number1
Product Code LWR
UDI-Device Identifier05415067033123
UDI-Public(01)05415067033123(17)270510(10)BR00038411
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberESP200-21
Device Lot NumberBR00038411
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2023
Initial Date FDA Received11/27/2023
Supplement Dates Manufacturer Received04/01/2024
Supplement Dates FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-