Model Number 865350 |
Device Problems
Device Alarm System (1012); No Audible Prompt/Feedback (2282)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/31/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported a speaker malfunction with the system.The device was not in use on a patient at the time of event, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Manufacturer Narrative
|
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of the evaluation could not confirm the customer's alleged malfunction.Per customer, it was indicated that there was speaker malfunction inoperative message at time of event.While at bench, the speaker produced audible sound.The speaker was confirmed to be functioning per specification.For precaution, the speaker was replaced per current process.The device was operational and returned to the customer.
|
|
Event Description
|
Philips received a complaint on the mx40 1.4 ghz smart hopping indicating that the unit had a speaker inoperative message with no tone.The device was not in use on a patient at the time of event, there was no adverse event reported.
|
|