Catalog Number XC200
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
It was reported a patient underwent a laparoscopic kidney and adrenal gland procedure on (b)(6) 2023 and absorbable clip was used.The clip could not be closed.Another device was used to complete the case.There were no adverse consequences to the patient.No device will be returning.Additional information has been requested.
 
Manufacturer Narrative
Product complaint # : (b)(4).Additional information: h6 component code: g07002 - device not returned.Additional information has been requested however not received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Package lot number of the clips? the lot number of the clips is td2aej.Please provide the applier product code and lot number? the product code is ka200 and lot number is unknown.Please confirm if there is an issue with the applier? no, there is no issue with ka200.If yes, please create a product complaint and provide the respective reference number(s).=n/a.What suture type and size was used? it was 2-0 vicryle.When the event occurred, was the suture placed near the hinge of the clip? yes.Were you able to lock the clip closed on the suture? no.They could load the clip into the applier.They set the thread on the tip of the applier, but could not lock and the clip was opened.Even though they squeezed the handle of the applier tightly, they could not fix it.If yes, after it closed, was the clip holding securely fixed on the suture? n/a.Was the applier checked for damaged (jaws straight and aligned)? currently unknown.If the clip did not close/hold on the suture, was the clip used in an application where the suture was under tension?=.Currently unknown.Please provide the lot number: td2aej.Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.We regularly contact with sale rep about the device returning.Note: events reported via: mw# 2210968-2023-09233, mw# 2210968-2023-09234, mw# 2210968-2023-09235.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 12/28/2023.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRA-TY ABSORBABLE X6::SUTURE CLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN ANGELO
3348 pulliam st
san angelo TX 76905
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18217181
MDR Text Key329130442
Report Number2210968-2023-09236
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705031206885
UDI-Public10705031206885
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K931492
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberXC200
Device Lot NumberTD2AEJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-