ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; NO MATCH Back to Search Results
Model Number CDDRA500Q
Device Problem Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
During an initial implant procedure on 20 nov 2023, it was reported that the header of the implantable cardioverter defibrillator (icd) was cloudy.The icd functioned appropriately when tested, but out of caution, the physician elected to implant a new icd.The patient was stable throughout the procedure.
 
Manufacturer Narrative
The reported event of header discoloration was confirmed.Interrogation of the device revealed the device was above eri when received.The cause of the reported event was due to molten polymer crystallization that occurs during the molding process of the resin used to manufacture the headers, and due to the gallant header¿s inherent volume and shape, cooling of the insulative material happens at a varied rate, and this results in visible organization observed as clouds or streaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18217327
MDR Text Key329131891
Report Number2017865-2023-93306
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Lot NumberA000144084
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/20/2023
Initial Date FDA Received11/28/2023
Supplement Dates Manufacturer Received01/26/2024
Supplement Dates FDA Received02/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
-
-