Model Number CDDRA500Q |
Device Problem
Product Quality Problem (1506)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/20/2023 |
Event Type
malfunction
|
Event Description
|
During an initial implant procedure on 20 nov 2023, it was reported that the header of the implantable cardioverter defibrillator (icd) was cloudy.The icd functioned appropriately when tested, but out of caution, the physician elected to implant a new icd.The patient was stable throughout the procedure.
|
|
Manufacturer Narrative
|
The reported event of header discoloration was confirmed.Interrogation of the device revealed the device was above eri when received.The cause of the reported event was due to molten polymer crystallization that occurs during the molding process of the resin used to manufacture the headers, and due to the gallant header¿s inherent volume and shape, cooling of the insulative material happens at a varied rate, and this results in visible organization observed as clouds or streaks.
|
|