DEPUY ORTHOPAEDICS INC US DXTEND GLENOSPHERE ECC D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE Back to Search Results
Catalog Number 130760042
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Joint Dislocation (2374)
Event Date 11/07/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Subject id: (b)(6).Study no: dots.Clinical adverse event received for dislocation of the reversed shoulder prosthesis.Event is serious and is considered moderate.Event is possibly related to device.Event is not related to procedure.Date of implant: (b)(6) 2023.Date of event: 07 nov 2023.(right shoulder).Treatment: closed reduction/manipulation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND GLENOSPHERE ECC D42MM
Type of Device
DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18220597
MDR Text Key329158561
Report Number1818910-2023-24135
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027751
UDI-Public10603295027751
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130760042
Device Lot Number5405585
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND HIGHMO PE CUP D42 +3MM; DXTEND SCREW LOCK D4.5X36MM; DXTEND SCREW LOCK D4.5X42MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND SCREW NO LOCK D4.5X24MM; GLOBAL UNITE FX RSA EPI 1; GLOBAL UNITE STD STEM SZ 12; STD CTR SCREW METAGLENE 10MM
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient SexMale
Patient Weight83 KG
-
-