DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/15/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during inspection, the cardiosave intra-aortic balloon pump (iabp) units system will not boot up.There was no patient involvement.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) was having system won't bootup and was not operational.There was no patient involvement.A getinge field service engineer (fse) evaluated the unit and determined that due to 8 years long life executive pcb is not working and fse replaced, executive pcb.Fse then found the power cord to be not retracting and replaced, ac power cord.Fse then found the display locks up and fse resolved the issue by replacing kit, display monitor to video gen board.Fse then completed full preventive maintenance and then tested the safety and functionality as per factory specifications and iabp was then returned to customer for clinical use.
|
|