The investigation reviewed the calibration data; the results were within specifications.The investigation reviewed the qc data; the qc was out of specified ranges (>+3 standard deviations-sd) before the patient sample was run.Product labeling states "the control intervals and limits should be adapted to each laboratory¿s individual requirements.Values obtained should all within the defined limits.Each laboratory should establish corrective measures to be taken if values fall outside the defined limits.If necessary, repeat the measurement of the samples concerned." the investigation reviewed the alarm trace; the trace had one "sample short" (b)(6)2023, 11:39:05); two "abnormal sample aspiration" ((b)(6)2023, 8:59:45 am and 08:11:49 am); and one "sample clot detection" ((b)(6)2023, t08:10:52) alarms.The customer did not use rack adapters used for the 13 mm primary tube.Product labeling states "warning! incorrect results may occur if tubes are not aligned vertically incorrect placement of tubes on racks may cause incorrect pipetting, which may lead to false results.Ensure that tubes and cups are always placed vertically and are inserted fully in the racks.Use the cup adapter for tubes with an outer diameter of 13 mm or less.Alternatively, use racks without stabilizers for 13 mm tubes." based on the provided data, a general reagent issue can most likely be excluded; another reagent kit of the same lot performed according to specifications at this customer site.The investigation determined the event was caused by the customer's failure to follow instructions when patient samples were run despite the qc being out of the specified ranges.The investigation did not identify a product problem.
|