ABBOTT MEDICAL MASTERS SERIES MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 25MJ-501
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on 05 november 2023, a 25mm masters series mechanical heart valve was chosen for implantation for a mitral valve replacement.When the valve was implanted, it opened and closed abnormally.It was decided to remove the valve and replace with another 25mm masters valve.The patient remained on cardiopulmonary bypass throughout the procedure and remained hemodynamically stable.Due to an extended 15 minutes to replace the valve, additional anesthetic medication was provided to the patient.There were no patient effects, and the patient was reported to be stable.No additional information was provided.
 
Manufacturer Narrative
An event of a valve opening and closing abnormally after implantation was reported.The valve was returned to abbott for analysis and the investigation found that both leaflets were able to fully open and close with no resistance occurring.Leaflet function was considered normal with both leaflets opening while having no asynchronous motion or leaflet malfunction.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18230764
MDR Text Key329278144
Report Number2135147-2023-05239
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number25MJ-501
Device Lot NumberC00001471
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-