Product complaint (b)(4).Investigation summary cssd reported.No information on extent of damage.Surgeon unknown.Part of consigned sets, trays and serial numbers unknown.Tagged as damaged and available for collection.Urgent replacement please for high volume hospital.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed that attune spacer block has all four posts chipped, fragments were not returned for evaluation.Additionally, the devices exhibits an overall appearance of heavy usage.No other defects were observed.The observed condition of the device was consistent with use of excessive force in a prying motion while trialing.Properly handling and attention to the approved use of the device diminishes the risk for this type of failure.The overall complaint was confirmed as the observed condition of the attune spacer block would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|