Catalog Number 6903190 |
Device Problems
Fracture (1260); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 03/2024) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during a dialysis catheter placement procedure via the right internal jugular vein, the guidewire was allegedly bent in the blood vessels.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported guidewire deformation and fracture as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, d4 (expiration date: 03/2024), g3, h6 (device).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a dialysis catheter placement procedure via the right internal jugular vein, the guidewire was allegedly bent in the blood vessels.It was further reported that the avulsion sheath was allegedly broken and the broken sheath was removed.There was no reported patient injury.
|
|