Model Number MI1200 SYNCHRONY |
Device Problems
Improper or Incorrect Procedure or Method (2017); Migration (4003)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 11/22/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
It was reported that the device was explanted due to the migration of the electrode.The user has been re-implanted.
|
|
Manufacturer Narrative
|
Conclsuion: device investigations did not reveal any device defect or problem which is expected to have been present whilst implanted.This finding was expected, because according to the recipient report the device was explanted due to the electrode being pulled out of cochlea during an ear cleaning procedure.Mechanical damages found during investigation are attributable to the removal surgery.This is a final report.
|
|
Event Description
|
The user did not have any speech understanding anymore.It was initially reported that the device was explanted due to the migration of the electrode.As per additional information, the electrode array was pulled out of the cochlear during an ear cleaning procedure.During the explantation surgery it was confirmed that the electrode array was completely out of the cochlea.A complete insertion during the initial surgery were confirmed.The user has been re-implanted.
|
|