A facility reported that while positioning the patient in the mayfield modified skull clamp (a1059), the double pin side slipped, resulting in a laceration to the head.The incident occurred while the patient was in a supine position with the head turned, causing a 2-3cm laceration to the right forehead.The scalp was irrigated and staples were used for closure.There was a delay of approximately 30 minutes.
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 the mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device found no device deficiencies that would have contributed to the reported complaint as slippage could not be duplicated.However, the investigation showed that the starbursts were worn, the index knob tight, and there was no movement of the swivel base in the locked condition.The torque settings of the index knob will be set during repair.On disassembly, a lot of chemical deposits were observed in the locking mechanism; the unit requires preventative maintenance and replacement of parts.Root cause - the reported complaint issue of "slippage" is not confirmed.Probable root cause of the reported complaint is improper or suboptimal placement of the skull clamp on the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|