Catalog Number EL5ML |
Device Problem
Delivered as Unsterile Product (1421)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/02/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unknown procedure, pinhole was found on the tyvek.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
Manufacturer Narrative
|
(b)(4) date sent: 11/30/2023 d4: batch # a9dp2d this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 12/29/2023.D4: batch # a9dp2d.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the el5ml devices was returned inside it's original packaging.Upon visual inspection, it was observed that the tyvek from the packaging was damaged; it was noted to have a hole in the tyvek, the hole was noted to be from the outside in.The event reported was confirmed and it is related to improper use of the device.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|