• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/19/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) pumping did not stop, but the screen froze, and the waveform disappeared.The iabp was switched to another cardiosave, and treatment resumed.There was no patient harm reported.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse was not able to reproduce the issue, but were concerned that the operation panel might malfunction, so as a precaution, the video generator board and touch screen was replaced and returned the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18241138
MDR Text Key329811934
Report Number2249723-2023-05107
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2023
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
-
-