Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) pumping did not stop, but the screen froze, and the waveform disappeared.The iabp was switched to another cardiosave, and treatment resumed.There was no patient harm reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse was not able to reproduce the issue, but were concerned that the operation panel might malfunction, so as a precaution, the video generator board and touch screen was replaced and returned the device.
|
|