• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problems Low Audible Alarm (1016); Inaudible or Unclear Audible Prompt/Feedback (2283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer reported that the speaker volume was very low on the mx40 monitor.The device was not in use monitoring a patient at the time of the reported issue.No adverse patient or user event was reported.
 
Manufacturer Narrative
Philips received a complaint that the mx40 1.4 ghz smart hopping device speaker volume was very low.No patient impact was reported.The device was sent to philips bench for evaluation.The repair facility technician (rft) confirms that the speaker produced sound during testing.Based on the information available and the testing conducted, philips was unable to replicate the reported problem.The reported problem was not confirmed.Although the speaker was confirmed to be functioning per specification during testing as the unit produced sound, the speaker was replaced.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18245216
MDR Text Key329457987
Report Number1218950-2023-00902
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received12/01/2023
Supplement Dates Manufacturer Received12/05/2023
Supplement Dates FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-