Model Number DW ARTHROSCOPY FLUID MANAGEMENT DEV
Device Problems Pumping Problem (3016); Intermittent Loss of Power (4016)
Patient Problem Insufficient Information (4580)
Event Date 03/04/2021
Event Type  malfunction  
Manufacturer Narrative
The complaint is not confirmed.No pressure or flowrate discrepancies were observed.The unit passed all bench tests during evaluation.
 
Event Description
It was reported that the pump was functioning intermittently during an rcr case.The caller had no more information.
 
Manufacturer Narrative
The complaint is not confirmed.No pressure or flowrate discrepancies were observed.The unit passed all bench tests during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DW ARTHROSCOPY FLUID MANAGEMENT DEV
Type of Device
ARTHROSCOPE AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18245600
MDR Text Key329468652
Report Number1220246-2023-09032
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00888867039377
UDI-Public00888867039377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDW ARTHROSCOPY FLUID MANAGEMENT DEV
Device Catalogue NumberAR-6480
Device Lot Number10168797
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2021
Date Manufacturer Received03/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-