Catalog Number 88710100 |
Device Problems
Loose or Intermittent Connection (1371); Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).E1 telephone number: (b)(6).G2 foreign: belgium.Evaluation and investigation are in process.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
It was reported that prior to surgery on (b)(6) 2023, the screw doesn't fit well on a dermatome an.This resulted in the width plate being too loose.Follow-up provided the serial number of the device, 700727.There was no patient involvement, impact, or harm reported.Due diligence information complete.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the width plate was loose as an incorrect screw was used; however, the repair has not been completed due to awaiting material for repair.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to use error.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information available.
|
|