Device Problems
Separation Failure (2547); Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2023
Event Type
malfunction
Event Description
During the initial implant procedure, the physician experienced difficulty removing the stylet from the left ventricular (lv) lead.While attempting to remove the stylet, the connector pin came loose from the body of the lead.The lv lead was explanted and replaced to complete the implant procedure.The patient was in stable condition.
Manufacturer Narrative
The reported events of difficulty removing the stylet and lead damage were confirmed.As received, a partial proximal portion of the lead with stuck stylet was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.The polytetrafluoroethylene (ptfe) coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin with the crimp sleeve were found pulled out of the connector assembly stretching the inner coil consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead during the procedure.The cause of the reported events of difficulty removing the stylet and lead damage was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin to be pulled out through the connector assembly.