ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TRANSPAC¿ IV MONITORING KIT, DISPOSABLE TRANSDUCER, 3ML SQUEEZE FLUSH DEVICE, MA; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 886-42584-05 |
Device Problems
Break (1069); Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2023 |
Event Type
malfunction
|
Event Description
|
The event involved a transpac iv monitoring kit.The reporter stated that after the healthcare provider connected the transducer, the 3-way stopcock was found broken after 2 minutes of infusion causing a leak of normal saline solution.The kit was replaced, and surgery was completed.There was patient involvement but no harm was reported as a consequence of this event.
|
|
Manufacturer Narrative
|
The device is expected to be returned for evaluation, however, it is not yet received.
|
|
Manufacturer Narrative
|
One used device was returned for evaluation on(b)(6) 2023.The reported complaint of leakage was confirmed on the returned set.An image was provided by the customer showing the area of the defect.During visual inspection, the 48.5" pressure tubing was found separated from both the female and male luer.When microscopically examined, insufficient adhesive coverage was observed on both the ends of the pressure tubing.The separation of the bond was due to the pressure tubing being tacky.The probable cause of the pressure tubing separated from the male and female luer had occurred due to insufficient adhesive coverage on the tubing during assembly.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
|
Search Alerts/Recalls
|
|
|