Model Number AIA-900
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2023
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) conducted a site visit and was able to confirm the problem by review of the error log.The fse was unable to reproduce the reported problem due to the reported issue being intermittent.The fse ordered a new test cup pick up assembly kit.The fse returned to the customer site at a later date and installed the new test cup pick up assembly kit and performed all necessary alignments.The fse validated the analyzer by running quality control (qc) with results within acceptable range and instructed the customer to run the instrument to verify operation.The aia-900 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no other similar complaints found during the searched period.The aia-900 operator's manual under section 12 flags and error messages states the following: [2161] c.Transfer cup pickup failure cause: the cup sensor s063 failed to detect the cup after the cup was grasped.Action: please contact the tosoh local representatives.Check s063, the cup pickup position, and the cup pickup operation.The most probable cause was due to failure of the test cup pick up assembly.
 
Event Description
A customer reported 2161 cup transfer cup pickup failure error on the aia-900 analyzer.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
Correction to section h6: component code: from cup to carrier.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key18253660
MDR Text Key329571811
Report Number3004529019-2023-00474
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189283992
UDI-Public04560189283992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/03/2023
Supplement Dates Manufacturer Received11/09/2023
Supplement Dates FDA Received12/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-