Catalog Number EMAX2PLUS |
Device Problems
Material Frayed (1262); Fail-Safe Did Not Operate (4046)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition that the device ran in a locked position and exhibited vibration, identified during service and repair, was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|
|
Event Description
|
It was reported by japan that during service and evaluation, it was determined that the motor device had damaged flex circuit, worn bearing, was loose, ran in locked position, leaked air, produced excessive noise, exhibited vibration, illegible labeling, damaged cord and cracked/damaged housing.It was further determined that the device failed pretest for visual assessment, loctite assessment, no short circuit between phases and connector body, safety assessment, noise assessment and air pump assessment.It was noted in the service order that the device had damaged hose.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|