Model Number 06000-200 |
Device Problem
Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The braun/welch allyn thermoscan pro 6000 ear thermometer is indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment.The probe cover is used as a sanitary barrier between the infrared thermometer and the ear canal.To help ensure accurate temperature measurements, the sensor itself is warmed to a temperature close to that of the human body.When the braun/welch allyn thermoscan is placed in the ear, it continuously monitors the infrared energy until a temperature equilibrium has been reached and an accurate measurement can be taken.Hillrom/welch allyn¿s investigation of pro 6000 ¿hot tip¿ complaints confirmed that aggressive cleaning and decontamination practices can cause liquid ingress.Consequently, liquid ingress can adversely affect the temperature sensor causing the pro 6000 device to behave inconsistently and to overheat the speculum tip.It is believed that replication of the malfunction with the returned devices from the customer at the manufacturing site has not been possible as any fluid that may have ingressed had likely evaporated, therefore not showing the ¿hot tip¿ malfunction.Based on hillrom¿s ability to replicate the malfunction of a ¿hot tip¿ on new devices that can potentially go above the built in risk mitigations of a safety cut off that could potentially cause a more serious injury we have deemed this complaint to be reportable.
|
|
Event Description
|
The customer reported that the probe tip of their braun pro6000 overheated.There was no reported injury associated with this event.This incident was captured under hillrom complaint ref # (b)(4).
|
|
Event Description
|
The customer reported that the probe tip of their braun pro6000 overheated.There was no reported injury associated with this event.This incident was captured under hillrom complaint ref # (b)(4).
|
|
Manufacturer Narrative
|
Hillrom/baxter is submitting this supplemental report based on updated information received from the customer.The customer changed the reported allegation of probe tip overheating to can¿t hold charge.Hillrom/baxter does not consider this to be a reportable malfunction.Based on this information, no further action is required.
|
|
Search Alerts/Recalls
|