It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1004 indicative of a short circuit condition detected during shock delivery, additionally, the device also recorded a code 1006 indicative of high voltage during charge repeat.Analysis of the device was performed and it was found that the device exhibited low out of range shock impedance measurements.Furthermore, oversensing was observed, which led to three inappropriate shocks, the last one was the one suspected to have triggered the codes on the device.Replacement of the system was advised.It was decided to turn the therapy off and schedule a system replacement in the near future.At this time, this device remains in service.No further adverse patient effects were reported.
|