TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Event Description
A customer reported the aia-360 analyzer screen is frozen upon starting the analyzer.The frozen screen occurred last week and was resolved by restarting the analyzer multiple times.The customer attempted to restart the analyzer again, but the screen will not change from the start up session and unable to select any selections.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for cardiac troponin i (ctnl 2) and creatine kinase mb (ck-mb).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by restarting the analyzer and reviewing the error logs and found errors that could be related to the main board.The fse replaced the main board, yet the screen remains frozen.Fse also replaced the screen and could not reproduce the problem, due to the errors in the error log that was associated with the main board fse left the new main board in.In addition, the fse restarted the analyzer multiple times and did not get a frozen screen after replacing the screen.Fse repaired and validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the searched period.The most probable cause of the reported event was due to the failures of the main board and screen.
 
Manufacturer Narrative
The main board and display unit were returned to the instrument service center.A visual inspection was performed on the returned parts and no damage was observed.Functional testing was performed and the boot up was not successful, which indicates an electronic component failure due to expected or random component failure without any design or manufacturing issue.The most probable cause of the reported event was due to the faulty main board and display unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key18263853
MDR Text Key329707507
Report Number3004529019-2023-00475
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189282896
UDI-Public04560189282896
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-