Model Number AIA-360 IMMUNOASSAY ANALYZER |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/13/2023 |
Event Type
malfunction
|
Event Description
|
A customer reported the aia-360 analyzer screen is frozen upon starting the analyzer.The frozen screen occurred last week and was resolved by restarting the analyzer multiple times.The customer attempted to restart the analyzer again, but the screen will not change from the start up session and unable to select any selections.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for cardiac troponin i (ctnl 2) and creatine kinase mb (ck-mb).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Manufacturer Narrative
|
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by restarting the analyzer and reviewing the error logs and found errors that could be related to the main board.The fse replaced the main board, yet the screen remains frozen.Fse also replaced the screen and could not reproduce the problem, due to the errors in the error log that was associated with the main board fse left the new main board in.In addition, the fse restarted the analyzer multiple times and did not get a frozen screen after replacing the screen.Fse repaired and validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the searched period.The most probable cause of the reported event was due to the failures of the main board and screen.
|
|
Manufacturer Narrative
|
The main board and display unit were returned to the instrument service center.A visual inspection was performed on the returned parts and no damage was observed.Functional testing was performed and the boot up was not successful, which indicates an electronic component failure due to expected or random component failure without any design or manufacturing issue.The most probable cause of the reported event was due to the faulty main board and display unit.
|
|
Search Alerts/Recalls
|