It was reported that electrogram (egm) review was requested for this cardiac resynchronization therapy defibrillator (crt-d) system due to undersensing of ventricular flutter with inappropriate pacing.The device was programmed to vvi 80 beats per minute (bpm), and right ventricular (rv) sensitivity was programmed to 0.6mv.Technical services (ts) explained that automatic gain control (agc) is set to 8 times the sensing floor when pacing, and the signal should be visible before the next paced event.The field representative measured the signal, and it was 0.2-0.4mv.It was noted that the patient had a left ventricular assist device (lvad) implanted three months prior, and the device sensing was fine prior to lvad placement.Ts reviewed that it was not unusual for the lvad to cause amplitudes to decrease.Technical services (ts) reviewed troubleshooting options and programming considerations.Anti-tachycardia pacing (atp) did not convert the rhythm, and commanded 41j shock was delivered.This crt-d system remains in service.No additional adverse patient effects were reported.
|