Model Number 72081-01 |
Device Problem
High Readings (2459)
|
Patient Problems
Hypoglycemia (1912); Tachycardia (2095); Dizziness (2194); Loss of consciousness (2418); Diaphoresis (2452)
|
Event Date 11/21/2023 |
Event Type
Injury
|
Event Description
|
A high reading issue was reported with the adc device.The customer reported receiving unspecified sensor scan results perceived to be high compared to unspecified readings on another adc meter.As a result, the customer experienced dizziness, sweating, palpitations, feeling of "lightheadedness" followed by loss of consciousness.The customer was able to self-treat with glucose and there were no reports of the customer requiring treatment by a healthcare processional (hcp) or third-party.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high reading issue was reported with the adc device.The customer reported receiving unspecified sensor scan results perceived to be high compared to unspecified readings on another adc meter.As a result, the customer experienced dizziness, sweating, palpitations, feeling of "lightheadedness" followed by loss of consciousness.The customer was able to self-treat with glucose and there were no reports of the customer requiring treatment by a healthcare processional (hcp) or third-party.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.Visual inspection performed on the returned sensor and no issue was observed.Data was extracted using approved software, and extraction was successful.The sensor was found to be in sensor state 6 (indicating early termination).The watermark was observed at the base of the tail indicating the sensor being properly inserted.The returned sensor was further investigated and de-cased.Performed an internal visual inspection on the sensor¿s pcba (printed circuit board assembly); no issues were observed.Performed an smu (source measurement unit) test with connector key to ensure the sensor's electronics were functioning correctly, and the returned unit did not have any glucose reading issues.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.No malfunction or product deficiency was identified.Therefore, this issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.This serves as a correction report: section(s) updated as follows: b1: "product problem" was incorrectly documented in previously submitted mdr and has been updated to "adverse event".
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.Visual inspection performed on the returned sensor and no issue was observed.Data was extracted using approved software, and extraction was successful.The sensor was found to be in sensor state 6 (indicating early termination).The watermark was observed at the base of the tail indicating the sensor being properly inserted.The returned sensor was further investigated and de-cased.Performed an internal visual inspection on the sensor¿s pcba (printed circuit board assembly); no issues were observed.Performed an smu (source measurement unit) test with connector key to ensure the sensor's electronics were functioning correctly, and the returned unit did not have any glucose reading issues.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.No malfunction or product deficiency was identified.Therefore, this issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high reading issue was reported with the adc device.The customer reported receiving unspecified sensor scan results perceived to be high compared to unspecified readings on another adc meter.As a result, the customer experienced dizziness, sweating, palpitations, feeling of "lightheadedness" followed by loss of consciousness.The customer was able to self-treat with glucose and there were no reports of the customer requiring treatment by a healthcare processional (hcp) or third-party.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Event Description
|
A high reading issue was reported with the adc device.The customer reported receiving unspecified sensor scan results perceived to be high compared to unspecified readings on another adc meter.As a result, the customer experienced dizziness, sweating, palpitations, feeling of "lightheadedness" followed by loss of consciousness.The customer was able to self-treat with glucose and there were no reports of the customer requiring treatment by a healthcare processional (hcp) or third-party.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.Visual inspection performed on the returned sensor and no issue was observed.Data was extracted using approved software, and extraction was successful.The sensor was found to be in sensor state 6 (indicating early termination).Sensor state 6 with event logs 6 and 7 are an indication that the patch was terminated without any error.The event logs observed are consistent with normal termination and indicate the sensor has reached its end of life.The watermark was observed at the base of the tail indicating the sensor being properly inserted.The returned sensor was further investigated and de-cased.Performed an internal visual inspection on the sensor¿s pcba (printed circuit board assembly); no issues were observed.Performed an smu (source measurement unit) test with connector key to ensure the sensor's electronics were functioning correctly, and the returned unit did not have any glucose reading issues.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.The passing of all functionality test indicates that there were no issues with sensor functionality and electronics.Therefore, this issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Search Alerts/Recalls
|
|