• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARJOHUNTLEIGH POLSKA SP. Z O.O. ARJO FLUSHER DETERGENT; WASHER OF BODY WASTE RECEPTACLES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJOHUNTLEIGH POLSKA SP. Z O.O. ARJO FLUSHER DETERGENT; WASHER OF BODY WASTE RECEPTACLES Back to Search Results
Device Problem Use of Device Problem (1670)
Patient Problem Sore Throat (2396)
Event Date 10/10/2023
Event Type  Injury  
Event Description
Arjo was informed of an event involving arjo liquid flusher rinse and arjo liquid flusher detergent used with some other liquids (i.E.Clements chemicals - non arjo products) to clean the waste transfer station area.After washing the floor, the waste transfer station was closed overnight (without ventilation) and opened early in the morning by transport workers.Transport workers complained about sore eyes/throat.Additional information was received that some workers were given eye drops and ointment.Some of them required another visit to the eye clinic and some required care in the hospital.
 
Manufacturer Narrative
This report refers to arjo liquid flusher detergent.The report for arjo liquid flusher rinse was registered separately (under the internal record tw(b)(4)).Based on the information collected in the course of the investigation, the most likely cause of the event might be related to use error - product labeling not being followed (arjo liquid flusher detergent) intended for automatic rinsing of human waste containers was used to clean the floor.This is in line with the customer statement included in the customer report that the product was misused by trust employees.It should be underlined that the flushers are equipped with a dosing system (pump and hoses) and the dosage of arjo liquids is controlled by a microprocessor in the flusher.Following the flushers instruction for use (ifu) (e.G.For amigo flusher 60015000202_4en) : the arjo liquid flusher detergent is: "a mild alkaline detergent to be used for automated cleaning of human waste containers." "exercise caution when handling the chemical agent used in the machine.Follow the manufacturers' instructions or contact the manufacturer: if the agent comes into contact with the eyes or skin or if vapors are inhalated".In summary, the device (arjo liquid flusher detergent) was up to the manufacturer¿s specifications during the event.The product was not used for a patient hygiene at the time of the event.This complaint was decided to be reported to the competent authorities due to indication that staff members sustained serious injuries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARJO FLUSHER DETERGENT
Type of Device
WASHER OF BODY WASTE RECEPTACLES
Manufacturer (Section D)
ARJOHUNTLEIGH POLSKA SP. Z O.O.
ul. ks. piotra wawrzyniaka 2
komorniki PL-62 052
PL  PL-62052
Manufacturer (Section G)
ARJOHUNTLEIGH POLSKA SP. Z O.O.
ul. ks. piotra wawrzyniaka 2
komorniki PL-62 052
PL   PL-62052
Manufacturer Contact
katarzyna bobrow
ks. wawrzyniaka 2
komorniki 62-05-2
PL   62-052
668046472
MDR Report Key18265734
MDR Text Key329719084
Report Number3007420694-2023-00292
Device Sequence Number1
Product Code FLH
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/06/2023
Initial Date FDA Received12/05/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-