• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER AND LOPEZ VALVE®; NASOGASTRIC TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER AND LOPEZ VALVE®; NASOGASTRIC TUBE Back to Search Results
Catalog Number 0056160
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that a crack found when inserting nasogastric tube into patient.They opened a second tube and the same issue.Pulled one with a different lot number and no problem.
 
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.Visual evaluation of the returned sample noted four unopened (without original packaging), nasogastric sump tube.Visual inspection of the sample noted the four nasogastric pump tube was observed and no crack or holes noted.No root cause could be found because the reported event was unconfirmed.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.As the reported event is unconfirmed a labeling review is not required.Correction: d, h h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that a crack found when inserting nasogastric tube into patient.They opened a second tube and the same issue.Pulled one with a different lot number and no problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER AND LOPEZ VALVE®
Type of Device
NASOGASTRIC TUBE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18265881
MDR Text Key329720791
Report Number1018233-2023-08729
Device Sequence Number1
Product Code FEG
UDI-Device Identifier00801741052392
UDI-Public(01)00801741052392
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0056160
Device Lot NumberNGHR0116
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received12/05/2023
Supplement Dates Manufacturer Received03/15/2024
Supplement Dates FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-