This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the likely cause of the reported event is due to the following: 1) during the output activation in seal & cut mode, the probe was contacting hard tissue, metal objects or surgical instruments.2) due to ultrasonic vibration, the coating of the probe peeled off.3) a force to activate the output in seal & cut mode, or a force to grasp the tissue was applied to the probe.Therefore, cracks were generated at the scratched area.4) a force was applied to the probe causing it to break.Additionally, wear of the tissue pad occurs due to the following mechanism due to the seal and cut was output when nothing was gripped by the gripper (including after the tissue was cut), the tissue pad was worn out and a portion of the tissue was torn.The coating of the grasping section could have come off when dirt such as burn was scraped off with something hard.The event can be detected and prevented by following the instructions for use (ifu) which state: ¿ the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.¿ do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.·do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ if the grasping section, metal-exposed area around it or the probe tip gets sticked tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.Olympus will continue to monitor field performance for this device.
|