D10 concomitant product: lf1837, blunt tip sealer divider lf1837 (lot#30790001x); vlft10gen, vlft10gen ft series energy platformx1 (serial# (b)(6)); vlft10gen, vlft10gen ft series energy platformx1 (serial# (b)(6)) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
According to the reporter, during a total laparoscopic hysterectomy procedure, while being used on omentum adhesions <(><<)>2-3mm thick, two different lf1837 from the same lot were providing seal tones quickly (.5-1 seconds) but seal was not complete/no evidence of energy being deployed.Both handpieces were attempted to be used on two different generators (ft10; sn: (b)(6) <(>&<)> (b)(6)) with same results.Activation tone and "seal cycle complete" tone were noted.There was no re-grasp alert.An end tone was heard within 5-1s.Issue occurred immediately so cleaning of jaws was unnecessary.Lxmj37s was used as a sub with no issues.There was minimal bleeding that does not require transfusion.
|