Model Number AIA-900
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Event Description
A customer reported error message "2236 bf probe 2 suction failure" on the aia-900 analyzer.The customer was instructed by a technical support specialist (tss) to replace the bf tips and clean the surrounding area as well as check the waste tubing.The error persisted.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg), estradiol (e2), follicle stimulating hormone (fsh), luteinizing hormone (lh ii) and progesterone (prog ii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineer (fse) conducted a site visit and was able to confirm the problem by reviewing the error log.The fse ran bf wash and checked wash probes (s132) and (s133).The wash probe (s132) led was flickered during performing wash prime.The fse replaced wash probe (s132) and resolved the complaint.The fse verified the resolution by successfully running controls.No further action is required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the search period.The aia-900 operators manual under section 12: flags and error messages state the following: [2236] bf probe 2 suction failure cause: the overflow sensor (b)(6) detected liquid after a washer suction operation.A wu flag will be attached to the measurement result.Action: clean the wash probe.Check s133, the waste liquid solenoid valve sv171, the waste liquid tube, and the liquid pump lp173.The most probable cause of the reported event was due to faulty wash probe (s132).
 
Manufacturer Narrative
Correction section d9: was the device returned to manufacturer? part was returned to isc per the tracking information provided, however, the part could not be located and a product evaluation could not be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key18271757
MDR Text Key329808191
Report Number3004529019-2023-00478
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-