It was reported that during a vena cava filter placement procedure via the left femoral vein, the filter allegedly got stuck during the process of pushing the filter into the guide sheath.It was further reported that the filter could not be pushed to the far end of the guide sheath normally, and allegedly failed after several attempts.The procedure was completed using another device.There was no reported patient injury.
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the denali filter products that are cleared in the us.The pro code and 510k number for the denali filter products is identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one denali femoral delivery system was returned for evaluation.Six photos were provided and reviewed.The photo shows an empty filter storage tube and loads into the introducer sheath and partial view of pusher catheter was observed and pusher wire noted to be detached.Visual evaluation of the returned device pusher wire noted to be detached and filter was observed within the introducer sheath.Bunching was noted on the distal portion of the introducer sheath.Based on the findings the investigation is confirmed for the reported failure to advance issue as the filter was in the introducer sheath.The investigation is also confirmed for the identified detached issue as the pusher wire was detached and the investigation is confirmed for the identified material deformation issue as bunching was noted on the distal portion of the introducer sheath.A definitive root cause for the reported failure to advance and identified detached and material deformation issues could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|