• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER Back to Search Results
Catalog Number DL950F
Device Problems Failure to Advance (2524); Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2023
Event Type  malfunction  
Event Description
It was reported that during a vena cava filter placement procedure via the left femoral vein, the filter allegedly got stuck during the process of pushing the filter into the guide sheath.It was further reported that the filter could not be pushed to the far end of the guide sheath normally, and allegedly failed after several attempts.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the denali filter products that are cleared in the us.The pro code and 510k number for the denali filter products is identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one denali femoral delivery system was returned for evaluation.Six photos were provided and reviewed.The photo shows an empty filter storage tube and loads into the introducer sheath and partial view of pusher catheter was observed and pusher wire noted to be detached.Visual evaluation of the returned device pusher wire noted to be detached and filter was observed within the introducer sheath.Bunching was noted on the distal portion of the introducer sheath.Based on the findings the investigation is confirmed for the reported failure to advance issue as the filter was in the introducer sheath.The investigation is also confirmed for the identified detached issue as the pusher wire was detached and the investigation is confirmed for the identified material deformation issue as bunching was noted on the distal portion of the introducer sheath.A definitive root cause for the reported failure to advance and identified detached and material deformation issues could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI FEMORAL SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18272563
MDR Text Key329809500
Report Number2020394-2023-01141
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDL950F
Device Lot NumberGFGY0244
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-