• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT Back to Search Results
Catalog Number 778426
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2023
Event Type  malfunction  
Event Description
It was reported that when the customer was using the ureteral stent, they took out the packaging box and found that the stent was broken.And a new stent was replaced and the surgery was continued.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted received one photo sample which shows top overview of stent and push catheter within enclosed bag with pigtail removed from the rest of the stent.Root cause could not be identified.Although a root cause could not be definitively identified, a potential root cause for this type of failure could be user does not handle with care, bends sharply.However, there was insufficient information to confirm this potential root cause.A dhr review did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that when the customer was using the ureteral stent, they took out the packaging box and found that the stent was broken.And a new stent was replaced and the surgery was continued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT
Type of Device
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18273725
MDR Text Key329828205
Report Number1018233-2023-08748
Device Sequence Number1
Product Code FAD
UDI-Device Identifier00801741014765
UDI-Public(01)00801741014765
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K983498
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number778426
Device Lot NumberNGHN1175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-