Model Number M4735A |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported to philips that the heart start xl does not hold a charge.Patient involvement is unknown at this time.However, no adverse patient impact or injury was reported by the customer.Additional information has been requested.
|
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
|
Manufacturer Narrative
|
This report is based on information provided by philips field service personnel and has been investigated by the philips complaint handling team.Philips received a complaint on the heartstart xl defibrillator/monitor indicating that the battery does not hold a charge.The device was not in clinical use at the time the issue was discovered.There was no reported patient impact / injury.Remote service engineer confirmed by contacting the customer.The functional check highlights the exhausted battery which is not charging correctly.The requested battery can no longer be ordered due to the obsolescence of the defibrillator.End of support letter was sent to the customer.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.H3 other text : remote support provided.
|
|