Model Number M3535A
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
It was reported to philips that the battery will not hold a charge.There was no reported patient impact or injury.
 
Manufacturer Narrative
This report is based on information provided by philips remote service personnel and has been investigated by the philips complaint handling team.Philips received a complaint on the heartstart mrx defibrillator indicating that the device does not hold the battery.There was no patient harm.Based on the information available, the root cause of reported issue is due to the battery being out of calibration.The device is working fine as per manufacturer's specifications after battery calibration was performed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key18275575
MDR Text Key329844765
Report Number3030677-2023-04816
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2023
Initial Date FDA Received12/06/2023
Supplement Dates Manufacturer Received11/29/2023
Supplement Dates FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-