Model Number 80014-002
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
Service error code was received on the customer support helpline queue.There was no patient injury or adverse event.However, the service required event code indicates a wcd system issue that if it were to recur could result in the wcd failing to provide needed therapy.
 
Manufacturer Narrative
Device evaluation of the alleged malfunction required retrieval of the device from the patient; return to the contract manufacturer for decontamination, inspection and initial testing; followed by shipment to the manufacturer of record for failure investigation and root cause analysis.Returned therapy cable passed weight, safety, and eit.The returned monitor was tested and passed both isl (safety) and eit (performance) testing.The returned device passed all field return tests and inspections.The reported condition was not able to be replicated.There is no indication of a product malfunction.The reported service error code alert can be attributed to a system power-on self-test fault detected when a patient plugs and unplugs the therapy cable during system boot up.Will continue to monitor and trend service code issues for failure modes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURE WEARABLE CARDIOVERTER DEFIBRILLATOR
Type of Device
WEARABLE AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
KESTRA MEDICAL TECHNOLOGIES, INC.
3933 lake washington blvd ne
ste 200
kirkland WA 98033
Manufacturer (Section G)
KESTRA MEDICAL TECHNOLOGIES, INC.
3933 lake washington blvd ne
ste 200
kirkland WA 98033
Manufacturer Contact
jay wiese
3933 lake washington blvd ne
ste 200
kirkland, WA 98033
2069701994
MDR Report Key18276834
MDR Text Key329854882
Report Number3015185344-2023-00243
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P200037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number80014-002
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/01/2023
Initial Date FDA Received12/06/2023
Supplement Dates Manufacturer Received12/01/2023
Supplement Dates FDA Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
Patient Weight284 KG
-
-