HOLOGIC, INC MYOSURE XL; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 50-501XL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Foreign Body In Patient (2687)
Event Date 11/20/2023
Event Type  Injury  
Event Description
It was reported that on (b)(6), a myosure procedure was performed and, during the case, the device made a crunching sound and it was noticed a piece of metal in the patient's cavity.The piece was removed with graspers and has been saved with the device.The doctor may have applied slight torque on the device a couple of times, but not much.The patient had a prior acessa and myosure procedure in (b)(6) 2023.Before that, she went to the er at another facility and had an embedded iud dealt with.No additional information available.
 
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE XL
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
MDR Report Key18278613
MDR Text Key329867305
Report Number1222780-2023-00456
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number50-501XL
Device Catalogue Number50-501XL
Device Lot Number23G13RF
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/20/2023
Initial Date FDA Received12/06/2023
Supplement Dates Manufacturer Received11/20/2023
Supplement Dates FDA Received01/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient SexFemale
-
-