Catalog Number CH2000S-C
Device Problem Fluid/Blood Leak (1250)
Patient Problems Chemical Exposure (2570); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Event Description
The event involved a spinning spiros® closed male luer, red cap.It was reported that there was leaking between spiros and tubing, chemo dripped on patient.There was patient involvement and no harm was reported.
 
Manufacturer Narrative
One (1) used.List #ch2000s-c, spinning spiros® connected with an used (b)(4) sodium chloride injection usp baxter 25 ml intravenous bag with chemotherapy and an used unknown infusion set were returned for evaluation.As received an unknown chemotherapy residuals and no damage on the spiro's splines and a good shear tab activation were observed.The set was primed and a leaks was confirmed between the spiro and the connection with the extension set.The residual torque was found within specification and male luer of the extension set and female luer of the spiro were within specification.Complaint of leaks can be confirmed based on the used phyical sample evaluation.The probable cause of the this event is unknown.A device history review could not be conducted because no lot number(s) was/were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINNING SPIROS® CLOSED MALE LUER, RED CAP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18279822
MDR Text Key329905551
Report Number9617594-2023-01085
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082806
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCH2000S-C
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/08/2023
Initial Date FDA Received12/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED CHEMOTHERAPY DRUG, MFR UNK
-
-