• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR Back to Search Results
Model Number 71005
Device Problem Malposition of Device (2616)
Patient Problem Inadequate Pain Relief (2388)
Event Date 11/11/2023
Event Type  Injury  
Manufacturer Narrative
During the initial implant procedure, the physician displayed difficulty tunneling to place the ipg due to patient's loose skin.The decision was made to place the ipg slightly deeper to avoid the skin issues, however this decision affected the functionality of the system.There are no indications that the nalu system or any components failed or malfunctioned.The communication issues appear to be directly related to the placement of the ipg during the implant procedure.
 
Event Description
Patient was implanted with the nalu spinal cord stimulator on (b)(6) 2023.Immediately after implant the patient reported poor communication between the implantable pulse generator (ipg) and the external therapy discs.Field investigation determined that the ipg was implanted beyond the recommended depth for proper communication.A surgical revision was performed on (b)(6)2023 in which a new ipg was placed in a more superficial pocket location to correct communication issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key18280215
MDR Text Key329888885
Report Number3015425075-2023-00305
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71005
Device Catalogue Number71005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-