Model Number 16-02-80
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that, then the heater-cooler system 3t has been turned on, there was a massive leak of water from the area above all 4 wheels.Additionally, water runs out of the ventilation.There was no patient involvement.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in germany.Cardioplegia block leaking water, patient pump 1 makes loud whining noises, bearing damage.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See intial report.
 
Manufacturer Narrative
The unit was beyond repairable, no repair activity was performed and the unit was put out of service.Analysis of complaints database resulted in no further similar leakage instances notified for this device since installation in 2017.Based on all the above facts, taking into account the evidences collected by technician and considering the manufacturing year of the device, it cannot be excluded that the most likely root cause of the reported event is related to the wear/material degradation of the unit due to wearing.The wearing of electro-mechanical devices can be originated by the specific use condition at customer site that may have contributed to the event, such as movements or liquids penetration.However, there is no concerning trend for this kind of failure.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3T HEATER COOLER
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18281081
MDR Text Key329895911
Report Number9611109-2023-00622
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/10/2023
Initial Date FDA Received12/07/2023
Supplement Dates Manufacturer Received02/16/2024
Supplement Dates FDA Received03/12/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-