DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problems
No Device Output (1435); Gas/Air Leak (2946); Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported during a routine check , the cardiosave intra-aortic balloon pump (iabp) was reported to have a helium leak, the rescue unit wheels were not opening and closing, and there was an occasional loss of signal from leads.There was no patient involvement reported.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate the issue.The fse troubleshot the problems and found no faults or errors.The fse tested the unit and no leaks were found or signal loss from leads.The rescue unit wheel assembly screws were loose.So, the fse tightened the screws as needed, adjusted and lubricated them.The wheels were then opening and closing correctly.The fse then let flight transporter test the rescue unit wheels, and all worked correctly.The fse made a suggestion to site contact and transporter to purchase new set of cables.The fse completed all functional and safety tests per manufacturer specifications.
|
|