DEPUY IRELAND - 9616671 ATTUNE RP CR ARTIC SURF SZ8; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Catalog Number 254500568
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2023
Event Type  malfunction  
Event Description
It was reported that the femoral impactor, tibial impactor and articulation surface are cracked.It did not happen in surgery.
 
Manufacturer Narrative
Product complaint # (b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: it was reported that the femoral impactor, tibial impactor and articulation surface are cracked.It did not happen in surgery.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned device found that the attune rp cr artic surf sz8 was broken from one of the post.Broken portion was nor returned additionally the device shows scratches and dents on it surface.No evidence of cracking were observed.No other issues were observed.The failure mode is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.The overall complaint was not confirmed as the observed condition of the attune rp cr artic surf sz8 would not have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP CR ARTIC SURF SZ8
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18282625
MDR Text Key329940484
Report Number1818910-2023-24842
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500568
Device Lot NumberMVMGBS880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2023
Initial Date FDA Received12/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-